EXPLORER II PTCA CATHETER

Design

»

Specifications

»

PUBLICATIONS

Stellium DES

»

S-Flex BMS

»

 

Explorer II
PTCA BALLOON CATHETER

Catheter Compliance

Pressure
(MPa / Bar)

Catheter Diameters (mm)

1.5 2.0 2.5 3.0 3.5 4.0
0.4 / 4.0 1.31 1.81 2.31 2.81 3.31 3.81
0.6 / 6 1.41 1.91 2.41 2.91 3.41 3.91
0.8 / 8 1.50 2.00 2.50 3.00 3.50 4.00
1.0 / 10 1.54 2.06 2.56 3.06 3.56 4.08
1.2 / 12 1.58 2.12 2.62 3.12 3.62 4.16
1.4 / 14 1.62 2.18 2.68 3.18 3.68 4.24
1.6 / 16 1.66 2.24 2.74 3.24 3.74 4.30
1.8 / 18 1.69 2.27 2.78 3.29 3.80 4.36
2.0 / 20 1.72 2.30 2.82 3.34 3.86 4.42
2.2 / 22 1.75 2.33 2.86 3.39 3.92 4.48
Ave Burst (Bar) 22 24 24 26 24 24

Do not exceed the Rated Burst Pressure, which is indicated in red.
 
Catheter Specifications
Catheter Length 135cm
Proximal shaft size 2.0F
Proximal shaft material Stain steel hypotube
Distal shaft material High density polyethylene
Recommended Guide Wire 0.014" (0.36mm) Max
Catheter Sizes and Order Codes
 

Balloon Diameter (mm)

Balloon length (mm)

1.5
2.0

2.5

3.0

3.5

4.0

10

EN2-1510
EN2-2010

EN2-2510

EN2-3010

EN2-3510

EN2-4010

12

EN2-1512
EN2-2012

EN2-2512

EN2-3012

EN2-3512

EN2-4012

15
EN2-1515
EN2-2015
EN2-2515
EN2-3015
EN2-3515
EN2-4015

20

EN2-1520
EN2-2020

EN2-2520

EN2-3020

EN2-3520

EN2-4020

25

EN2-1525
EN2-2025

EN2-2525

EN2-3025

EN2-3525

EN2-4025

30

-
EN2-2030

EN2-2530

EN2-3030

EN2-3530

EN2-4030

Indications for Use
Explorer II Percutaneous Transluminal Coronary Angioplasty (PTCA) Balloon Catheter is indicated for balloon dilatation for the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion.
Contra-indications
  • Unprotected left main coronary artery.
  • Coronary artery spasm in the absence of a significant stenosis.